ViroMax™ bacterial-viral filter features standard connectors (ISO 5356-1) for applications in breathing circuits, scavenging circuits, mechanical ventilation, and manual ventilation, including bag valve mask (BVM). The ViroMax filter is designed for single-patient, single-session use and comes in a case of 50 filters.
Low Flow Resistance: 1.5 cm H20 at 60 LPM
Coaxial ISO connectors:
22 mm inside diameter / 25 mm outside diameter
15 mm inside diameter / 22 mm outside
Low Flow Resistance: 1.5 cm H20 at 60 LPM
Compact, transparent housing
Bi-directional airflow
Disposable: for single-patient use only
Naturally "hydrophobic"
Ultrasonically welded
50 Filters per case
The ViroMax Viral-Bacterial Filter was tested at Nelson Laboratories at an increased challenge level that included a higher viral concentration and aerosolized particles that are smaller than the upper range (≤5 µm) of what the CDC generally defines as droplet nuclei 1 or infectious aerosols 2.
The virus used in these tests is the φX174, a spherical, icosahedral bacteriophage with 5,386 nucleotides and 27 nm (0.027 µm) in size. The challenge liquid was nebulized and then aerosolized in order to mimic aerosols and respiratory droplets. The mean particle size of these aerosols in the tests was 2.8 µm.
Each challenge in these reproducible tests consisted of 1.6 x 106 plaque-forming units or more. And the plates were incubated at body temperature for 12 to 24 hours.