High sensitivity measurement of Treponemal antibody in serum and plasma
Enables rapid syphilis testing with plasma sample. Also suitable for non-office hour and emergency testing
Provides quantitative test results by automated testing for objective diagnosis
Mixing unnecessary - Set up reagents and use without mixing
Use on various automated analyzers
Measurement Principle
IMMUNOTICLES AUTO3 TP is based on an optical absorbance method using Latex agglutination method.
Polystyrene-latex with TP(Treponemal) antigen* on the surface and TP antibody in serum form agglutination clump by antigen-antibody reaction. Based on the principle that photometric change by this agglutination reaction relies on TP antibody value, this reagent measures TP antibody value in samples from the calibration curve drawn using TP antibody calibrator.
*Treponemal Pallidum, Nichols strain
Absorption test
In order to confirm that the positive results obtained using this reagent is not a result of a non-specific reaction, Neutralization Reagent (sold separately) for IMMUNOTICLES AUTO3 TP is provided. With the presence of TP antibody in serum or plasma, a specific reaction occurs between TP antibody and TP antigen in the Neutralization Reagent when it is added. As a result, TP antibody titer is lower considerably in comparison to the value before absorption.
Measurement unit
QC pooled serum with the result of 1280 fold by TPHA quantitative test corresponds to the measurement result of 1280U with IMMUNOTICLES AUTO3 TP.
When WHO International Standard is measured, the result holds a relation of 1.0U=2.15mlU.