Measuring principles: Absorbance photometry Turbidimetry ,End-point,Two-point, Kinetic,/Dual/ reagent chemistries,monochromatic /bichromatic
8 step laundry system
Memory Upto 200,000 patient data
User friendly intellectual software
Independent mixing probe
TECHNICAL SPECIFICATION:
Reaction system
90 Special cuvettes
Length of cuvette – 6mm
Reaction time – 8 – 10 mins
Reaction temperature:37±0.1°c
Methodology
Measuring principles: Absorbance photometry, Turbidimetry ,End-point, Two-point, Kinetic,/Dual/ reagent chemistries,monochromatic /bichromatic
Power Requirements
Power supply AC100-240V, 50-60Hz
1 KVA
Sample Arrangement
Compatible with primary collection tube, minimal tube, sample cup, etc
Operating configuration
Operation system Windows 2000 or XP Interface RS-232
Memory Upto 200,000 patient data
Temperature 10°c-35°c Humidity upto 90% with no dew
Reagent arrangement
Reagent Probe – Liquid level detection, collision protection function
On board cooling ( 2°c- 8°c )
Laundry system
Needles: 8 step washing sequence.
Water Consumption- Upto 4 l/hr
Mixing system
Independent mixing probe
Optical system
9 wavelengths: 300-700nm
Tungsten halogen lamp
Absorbance range:0-4.00Abs
Spectrophotometry; rear spectrophotometry