Standard care for newborns including basic treatment in delivery rooms.
Stabilization of hypothermic conditions of newborns.
Treatment of hyperbilirubinemia of newborns with highly effective LED phototherapy.
Administering infusions and performing exanquinations.
Treatment of pathological newborns during controlled ventilation in the NICU.
Performing surgical procedures in the NICU and pediatric surgery departments.
Preoperative and postoperative intensive care in paediatric and neonatal wards.
Basic device configuration:
mobile stand with four traveling swivel casters (2 of them with brake)
upper heating module with 2 heating elements, built-in surface lighting and space for phototherapy installation
hydraulically infinitely adjustable bed (+/-22°), 650 x 800 mm, hinged plastic sidewalls
set of shelves: 3 adjustable for hanging on the stand, 1 fixed under the patient’s bed
adjustable infusion solution holder, 10 x 25 mm euro rail
electronic temperature controller module with manual and automatic proportional modes temperature control, self-diagnostics, 2 temperature sensors
touch keypad control, set of safety sound and light alarms
complete wiring for connection of other devices
Optional accessories:
heating pad HM-13
central oxygen distribution with 4 outlets
oxygen extractor with vacuum control
APGAR time module integrated in the central stand
halogen spotlight on flexible arm
plastic oxygen tent with a diameter of 500 mm
Technical parameters:
Instrument protection class: I, type BF
Supply voltage: 230/50 V/Hz
Power input: max. 1000 W
Power of the heating element: 600 W
Bed area LED lighting: 8/230 W/V
Spot LED lighting: 6/12 W/V