Light weight Anesthesia Work Station with LED display and single vaporizer
TECHNICAL SPECIFICATIONS
Tidal Volume 50ml - 1600ml
Ventilation Modes IPPV, Manual
Drive Modes Pneumatically driven, Electronically controlled
Gauges 3-Oxygen, 3-Nitrogen, 1-Air
Display LED display panel
Battery Backup 120 minutes
Measurement
Tidal Volume
Breathe frequency
MV
Alarms
MV High and MV low alarm
TV high and low alarm
Circuit disconnection alarm
O2 Fail safe
Hypoxic mixture fail safe
Optional
Two interlocked vaporizer
Poor fill
Basal flow
Gas Supply
O2, N2O, Air
Gas Pressure
O2 : 0.25Mpa ~ 0.65Mpa
N2O : 0.25Mpa ~ 0.65Mpa
Air : 0.25Mpa ~ 0.65Mpa
Flow Meter
Five tube flow meter Bobbin type
O2 tube : 0.1L/min ~ 1L/max, 1L/min 10L/max
N2O tube : 0.1L/min~1L/max, 1L/min 12/L/max
Air tube : 0.1L/min ~ 15L/min
O2 and N2O Linkage device, Hypoxy guard
Oxygen Flush
25L/min ~ 75L/min
Battery discharged, O2 supply features, Apnea, Circuit occlusion
Frame
Rust proof epoxy coated high strength frame / metal reinforced fiber frame (optional)
2 big drawers
Pipeline attachment for O2, N2O, Air
Yoke-2 each for O2, N2O
Vaporizer
Selectatec bar support, 2 interlocked vaporizers - Halothane / Isoflurane / Sevoflurane
Power Requirement
220V/110V, 50/60 Hz, Single Phase, Independent Earthing