AmcareMed medical LCD liquid oxygen alarm consists of a LCD display panel, control circuit and sensor. It is the monitoring part of the medical gas system.
The liquid oxygen alarm can monitor the liquid oxygen storage tank reserves, liquid oxygen tank pressure, oxygen temperature in the pipeline after vaporization, and oxygen pressure reduction station output pressure in real time. When the monitoring value is abnormal, it will sound and light alarm; it can connect 6 sensor signals.
Functions
* The visual alarm can be clearly identified at a distance of 4m, a visual angle of less than 30° and an illumination of 100lx.
* Each monitoring sampling point of the equipment has an independent alarm display, which lasts until the fault is resolved.
* The sound alarm means that the sound pressure level at 1.5m is not less than 55dB(A), and it has a temporary mute function.
* When the sound alarm sounds, the sound continues to alarm; if the sound off button is pressed, the sound alarm can be turned off.
* The sound alarm of the equipment is unconditionally started, which means that when an alarm is muted and other alarms occur, the sound alarm can be restarted.
* It has a visual alarm fault test function.
* It has a power-off recovery self-start function