AJ 5808-S INFUSION PUMP WITH SIMPLIFIED KEYPAD
Volumetric infusion pump delivers intravenous fluids and medication to patients in hospitals, outpatient surgical centers, hospices, nursing homes, and in ambulances. It is widely used throughout the hospitals and care centers including in the emergency room, ICU, CCU, labor and delivery, and in pediatric units.
Main features:
OLED display with improved brightness and contrast than LCD screens
Viewing the event log with time and date
Ability to operate the pump with IV sets or other marks designated
Two modes of infusion
Specifications
Infusion rate range
0.1-1200 ml/h (or up to 200 drops/min)
Infusion modes
• ml/h
• Drop/min
• Infusion time limit *
• Patient weight *
Volume to be infused
1 - 9999 ml
Bolus rate range
1 - 1200 ml/h
Bolus volume
1 - 9999 ml
Accuracy
± 2% on the pump only (± 5% with I.V set)
Occlusion sensitivity adjustment
High, Medium, Low, Lowest
0.04 MPa -0.16 MPa / 5.8 PSI ~23.2 PSI
KVO (Keep Vein Open)
0.1 ml/h – 3 ml/h (depends on the flow rate)
Built-in infusion rates memory
10 preset infusion modes memory
Built-in multi-languages display
English, Spanish, French, Italian
Alarm volume adjustment
High, Medium, Low
Alarms
• Door open
• Occlusion (up & down)
• Power source change
• Start reminder
• Infusion completed
• I.V set disengaged -
• Empty bag
• Initialization error
• Drive mechanism error
• Air-in line
• Battery low
• Battery depleted
Event log
400 + events
Safety protection
• Anti-free flow protection
• Automatic keypad lock
• Audible & visual alarms
• Password protection
Power supply
AC 100- 240V 50/60 Hz; DC 12V