600 and 1200 liters environmental chambers for Stability testing and controlled storage applications. Flexible interior allowing for different shelving configurations. Very precise and uniform stability incubators with state-of-the-art temperature and humidity control.
Meeting ICH Q1A guidelines for Stability Testing, these ‘reach-in’ stability cabinets have unsurpassed reliability in recreating Temperature and Humidity conditions. With FDA 21 CFR part 11 compliant data logging and IQ, OQ and PQ documents for mandatory qualifications.
PRODUCT DETAILS
Interior dimensions (H x W x D)
FitoClima 600:
1340 x 600 x 660 mm
52.76 x 23.62 x 25.98 inches
FitoClima 1200:
1340 x 1320 x 660 mm
52.76 x 51.97 x 25.98 inches
Exterior dimensions (H x W x D)
FitoClima 600:
1980 x 730 x 920 mm
78.0 x 28.74 x 36.22 inches
FitoClima 1200:
1980 x 1450 x 810 mm
78.0 x 57.09 x 31.89 inches
Temperature uniformity (ºc)
+/- 1ºc
Humidity range
20% to 95% rH
Humidity uniformity (%)
+/- 2% rH
Shelves
FitoClima 600: 4 shelves. Up to 10 can be fitted
FitoClima 1200: 8 shelves. Up to 20 can be fitted
Electrical connection
Single phase. Available with all regional electrical requirements