Cabinet Options:
• To embed in wall ABS Thermoplastic injection front meeting NFPA 286, galvanized steel sheet alarm mounting chasis made.
• To superimpose on a wall made of galvanized steel sheet white color powder coating finish.
Gas Number:
• To 1 and up to 8 simultaneous médical gases in the médical gas alarm zone.
Graphics:
• Pressure and vacuum readings on real time.
• NFPA99 color and Médical gas Name option.
• ISO color and Médical gas Name option.
• Flashing Alarms in each médical gas Module.
• Disconnections and Crossing of médical gas sensors.
• Display of area name where the médical gas alarm zone will be located.
Alarm Events:
• High and Low pressure Alarms on real time.
• Signal interruption or disconnection in the sensors Alarm.
• Cross connection of médical gas sensors Alarm.
• On-screen warning when the alarm activation parameters are not those specified by the NFPA99-2015 code
Communications:
• RD45 female ports for data input and output.
• Dry contact relay port for data output (1 port).
• RS485 communication protocol in SERIAL connection to connection with zone alarm groups and with Arigmed's SERIES 2 master alarm panel.
• Connection to INTRANET and EXTRANET (INTERNET) through the ARIGMED ETHERNET INTERFACE with HTML to view the system with remote devices such as PC, TABLET, SMARTPHONE via website.
Features:
• Parameter modifications by user to change Units, cut-off levels for pressure and Hysteresis.
• Editing Name of the area where the zone alarm will be located.