The OM-6060 measures the osmotic pressure of body fluids such as blood serum, plasma and urine.
Noise-free and accurate measurement using Freezing-point depression
Micro sample volume
Required sample volume: only 200 μL
Free calibration
3-point calibration
2-point calibration
Calibration data is memorized unless the power is turned off
Various measurement modes
Osmotic pressure ratio measurement
Control measurement
Specifications
Measurement objects - Blood serum, plasma, and urine *1
Measurement items - Osmotic pressure and osmotic pressure ratio of body fluid
Measurement range - 0 - 2000 mOsm (Switchable to "0 - 2500 mOsm"*2)
Measurement principle - Freezing-point depression using ultra supercooling
Sample supply - Turntable unit
Measurement precision - CV 1 % or less (200 - 300 mOsm)
Processing speed - 2 to 3 minutes per sample
Required sample volume - Minimum 200 μL (500-μL sample cup),
Minimum 2 mL (o.d. 12.3 × height 100 mm sample collection tube)
Sample container - 500-μL sample cup, sample collection tube (o.d.12.3 × height 100 mm)
Number of measurement samples - Maximum 24 samples
Calibration - 3-point calibration (0, 300, 1000 mOsm : polygonal line approximation),
2-point calibration (any 2 points : logarithmic curve approximation)
Display - 24-digit by 2-line LCD
Built-in printer - 58-mm width thermal printer paper (24 digits)
Memory capacity - 500 measurement results
External output - Compliant with RS-232C standard (Compatible with OM-6050, OM-6040, OM-6030, and OM-6020) Ethernet (option)
Measurement environment - Temperature: 10 - 30°C; Humidity: 20 - 80% RH (No condensation)
Dimensions - 320 (W) × 460 (D) × 447 (H) mm