Low albumin loss.
High performance membrane for various conditions of patients especially for elderly patients, hypoalbuminemia patients etc.
Gamma sterilized wet-type dialyzer.
Conditions:
1) Clearances: Qd=500mL/min, Qf=0mL/min
2) KoA: Qb=300mL/min, Qd=500mL/min, Qf=0mL/min
3) KUF: bovine blood, TP=60±5g/L, Hct=32±2%, Qb=300mL/min
4) Sieving Coefficient: bovine plasma, TP=60±5g/L
ISO 8637:2004
Specifications
Internal Diameter of Hollow Fiber (μm) 200
Wall Thickness of Hollow Fiber (μm) 45
Priming Volume (mL) 80/90/105/114
Maximum TMP (mmHg (kPa)) 500 (66)
Maximum Blood Flow (mL/min) 500
Maximum Dialysate Flow (mL/min) 800
Dimensions (mm[L] x mm[D]) 304 x 35/304 x 38/330 x 42/330 x 44
Weight (g) - 181 /205/486/510
Sterilization Gamma-Ray