Portable in size, available in both monopolar and bipolar modes, convenient handle power control and memory function, suitable for micro surgery, gynecology and arthroscopy.
Technical parameters:
Environmental temperature range:
5°C~ 40°C
Relative moisture range:
≤80%
Atmosphere pressure range:
86.0 ~ 106.0 Kpa
Working modes:
Pure cut:
1w ~ 60w (Load 500Ω)
Blend:
1w ~ 60w (Load 800Ω)
Coagulation:
1w ~ 60w (Load 1000Ω)
Fulguration:
0.1w ~ 35w (Load 1000Ω)
Biploar:
1w ~ 30w (Load 200Ω)
Technology data:
Voltage:
220V±10V, 50Hz ±1Hz
Working frequency:
357 KHz ~ 575 KHz
Leakage currents to ground:
Less than 0.5mA
Leakage currents to patient:
Less than 0.01mA
Display Setting:
Fulguration
a.0-10 range: 0.1watt increments
b.10-35range: 1.0 watt increments
Cut, Blend, Coagulation and Bipolar
1.0 watt increments
Power supply:220V±10%,50±1%Hz.
Transportation and storage
a)Ambient temperature range: -40°C~55°C;
b) Relative moisture range: ≤95%;
c)Atmosphere pressure range: 500hPa~1060hPa.
Key Features
OBS-100 Electrosurgical Generator with state-of-art technology offers superb performance on the electrco-surgery, and shows remarkable safety, reliability and convenience.
1 .Two levels of coagulation: pinpoint coagulation and fulguration
Pinpoint coagualtion provides précisé control of bleeding in localized areas.
Fulguration provides greater control of bleeding in highly vascular over surface areas.
2.Presets
The unit incorporâtes as many as ten user-defined presets for easy recall of frequently and settings.
3.lsolated RF output for Cut, Blend and Coag modes
This minimizes the potential of alternate site burns.