VETA - Series
Hyperbaric Oxygen Therapy Chamber System for Horses
Max. Working Pressure: 2.0 bar
Safety Valve Set Pressure: 2.1 bar
Design Pressure: 2.2 bar
Hydrostatic Test Pressure: 3.5 bar
Chamber Capacity (VETA 2000): 1 horse + 1 assistant/nurse
Chamber Capacity (VETA 3500): 2 horses + 1 assistant/nurse
Design and Manufacturing Certification: EN 13485
Design Standard: AD-2000 MERKBLATT
Pressure Chamber Certification: 2014/68/EU (PED CE)
Fire Suppression System Certification: NFPA 99
Electrical System Certification: EN 60601-1-2 EMC + EN 60601-1-2 LVD
Harmonized Standard: EN 14931
Manufacturing Period: 180 to 210 days
Commissioning and Training Period: 10 days
STANDARD EQUIPMENT
Manual Control System
Analyzing System
Emergency Safety System
Stainless Steel Tray System
OPTIONAL EQUIPMENT
Automation System
Inner Control System
Oxygen Supply System
Custom Exterior Livery
STANDARD SERVICES
Certification
Documentation
1-Year Warranty for Factory Defects
User Manual and Training
Preventative Maintenance Manual
Supervision on Installation
Commissioning
2-Year Online Technical Support
OPTIONAL SERVICES
Annual A-Maintenance
5-Year B-Maintenance
10-Year C-Maintenance
Unscheduled Maintenance
4-Year Extended Factory Warranty
Spare Parts and Equipment