PETA - Series
Hyperbaric Oxygen Treatment Chamber System for Pet Animals
Max. Working Pressure: 1.0 bar
Safety Valve Set Pressure: 1.1 bar
Design Pressure: 1.2 bar
Hydrostatic Test Pressure: 1.5 bar
Chamber Capacity (PETA 400): 1 pet animal
Chamber Capacity (PETA 600): 2 pet animal
Chamber Capacity (PETA 800): 3 pet animal
Chamber Capacity (PETA 1000): 4 pet animal
Design and Manufacturing Certification: EN 13485
Design Standard: AD-2000 MERKBLATT
Pressure Chamber Certification: 2014/68/EU (PED CE)
Medical Device Certification: 93/24/EEC
Fire Suppression System Certification: NFPA 99
Electrical System Certification: EN 60601-1-2 EMC + EN 60601-1-2 LVD
Harmonized Standard: EN 14931
Manufacturing Period: 30 to 45 days
Commissioning and Training Period: 5 days
STANDARD EQUIPMENT
Manual Control System
Analyzing System
Emergency Safety System
Stainless Steel Tray System
OPTIONAL EQUIPMENT
Automation System
Oxygen Supply System
Custom Exterior Livery
STANDARD SERVICES
Certification
Documentation
1-Year Warranty for Factory Defects
User Manual and Training
Preventative Maintenance Manual
Supervision on Installation
Commissioning
2-Year Online Technical Support
OPTIONAL SERVICES
Annual A-Maintenance
5-Year B-Maintenance
10-Year C-Maintenance
Unscheduled Maintenance
4-Year Extended Factory Warranty