Semi-automatic changeover
Contact gauges connected to each cylinder bank regulator will provide alarm notification when the duty bank is empty
Primary manifold control system for private facilities, small clinics, etc where local standards are applicable
Back up supply for medical air plant systems (HTM2022)
Flow capacity is 1200 lpm at 4 bar and 2000 lpm at 7 or 11 bar
The manual manifold is designed for use in medical or laboratory gas applications
Change over to the standby bank of cylinders is achieved by turning off the empty cylinders, and opening the standby cylinders for continuity of gas supply.
Flow capacity is 1200 lpm at 4 bar and 2000 lpm at 7 or 11 bar
Two cylinder emergency supply manifold complies to HTM 02-01 and ISO 7396-1
4 Bar, 7 Bar and 11 Bar pressures
Regulators comply with BS EN ISO 10524-2, test reports available on request
Test point (Gem 10) included as per code requirements
Supplied with integral 2 x 1 manifold header - extensions can be added for extra cylinder capacity
Flow capacity is 1200 lpm at 4 bar and 2000 lpm at 7 or 11 bar
Two cylinder emergency supply manifold complies to HTM 02-01 and ISO 7396-1
4 Bar, 7 Bar and 11 Bar pressures
Regulators comply with BS EN ISO 10524-2, test reports available on request
Test point included as per code requirements
Supplied with integral 2 x 1 manifold header - extensions can be added for extra cylinder capacity
Flow capacity is 200 lpm at 4 bar