The Area valve to be installed downstream of the line reducers as required by the standard for medical gas systems; a single device that meets all the regulatory indications of EN ISO 7396-1 relating to the Area Valve; consisting of a ball valve, inlet for specific gas maintenance, connection to the system with connection for Ø 16 mm welded pipe, physical disconnection, seat with check valve prepared for the transducer or the pressure switch for the mandatory signaling of pressure maximum and minimum of the line; predisposed for the installation of a proximity sensor, able to indicate the ON / OFF status of the area valve by transmitting the signal to the relative valve status alarm module.
N.One body made of CW617N chemically nickel-plated brass.
N.One Ø 16 mm inlet pipe connector.
N.One Ø 16 mm outlet pipe connector.
N.One inlet shut-off ball valve.
N.One emergency socket.
N.One physical disconnection.
NYLON shut-off valve gaskets.
FKM shut-off valve stem and O-Ring for Oxygen, NBR for other gases.
EN ISO 7396-1 | NF S 90-116 | UNI 9507 | EN ISO 5359 | DIN 13260-2 | BS 5682 | EN ISO 15001