Blender can provide the air flow with precise oxygen concentration to infant patients and effectively prevent the inhalation of high concentration of oxygen for a long time from causing damage to the central nervous system, respiratory system and retina of the brain.
Suitable for neonatal and obstetrics. Can be used with nasal oxygen catheters, baby incubators, oxygen masks, nebulizer, CPR and ECMO, etc.
Applicable users: Patients of premature infants, newborns, infants and children with body weight more than 1KG.
1.Treatment Time: 7*24H.
2.Continuously adjustable oxygen concentration flow, with precise control. 3.Low requirement of external air supply pressure and strong self-adaptability of pressure. 4.It can be used to cooperate with oxygen inhalation hood, nasal oxygen tube, warm box. for infants, cardiopulmonary resuscitator and others.
5.High Precision, good durability and stable output airflow.
Technical Security:
Air Flow: 0-15L/min,0-25L/min,0-30L/min, can be optional
Standard Configuration: Single Outlet,
Upgraded Configuration: Double Outlet
Oxygen Concentration: 21-100%
Air Supply Pressure: 0.3-0.5Mpa
Source: Air(medical use), Oxygen(medical use)
System Security:
Air/Oxygen Pressure Differential Alarm
Main Accessories:
Blender, Moisture bottles, Nasal Prong, High Pressure Air/Oxygen Tube