Features:
1. Accurately measure ETCO2, FiCO2, RR, SPO2, PR.
2. Equipped with body temperature and pulmonary pressure compensation (BTPS) and atmospheric pressure automatic compensation system, no manual intervention is required for zero calibration, and can be used in plateau environments and helicopter rescues.
3.3.5-inch touch screen operation, easy to learn, light weight, adjustable alarm volume, easy to carry and monitor.
4. The patented filter is protected from moisture, secretions, bacteria and dust. It has a small dead space and a large water storage capacity, and can work continuously in a humid environment (12h @50ml/min, 37°C, 100 % RH; 8h @ 100ml/min, 37°C, 100% RH).
5. Multiple data output modes: Upload real-time and historical data to PC via USB; transmit real-time data to mobile data terminal via Bluetooth; optional analog output for connecting to various polysomnography monitors.
6. Supports 8 languages.
Capnograph Measuring Data
EtCO2/CO2
Range: 0-150mmHg( 0-20kPa,or 0-20%(V/V))
Accuracy:
±2mmHg when 0-40mmHg,
±5%(reading) when > 41-70mmHg
±8%(reading) when > 71-100mmHg
±10%(reading) when > 101-150mmHg
EtCO2 Update: every breath,10sec,20sec or 30sec selectable
Trend length:24hours
Respiration rate
Range: 3 -150 t/min
Accuracy: the bigger value of ±1%(reading) or ±1times/min
Trend length: 24 hours
Flux : 50-250cc/min selectable
Oximeter Data (optional)
Oximetry Saturation
Range:0-100%
Accuracy:
±3% when 50-69%
±2% when 70-100%
Trend length: 24hours
Pulse Rate
Range:30-250 bpm
Accuracy:the bigger value of ±2bpm or ±2%(reading) when 30-250bpm