1. Designed for Small-Medium level Lab.
2. Rotational Cone plate method.
3. Non-Newtonian standard marker win China National Certification.
4. Original Non-Newtonian Controls, Consumables and application make whole solution.
Product Detail
Analyzer Introduction
SA-6000 automated blood rheology analyzer adopts cone/plate type measurement mode. The product imposes a controlled stress on the fluid to be measured through a low inertial torque motor. The drive shaft is maintained in the central position by a low resistance magnetic levitation bearing, which transfers the imposed stress to the fluid to be measured and whose measuring head is cone-plate type. The whole mensuration is automatically controlled by the computer. The shear rate can be set randomly at the range of (1~200) s-1, and can trace two-dimensional curve for shear rate and viscosity in real time. The measuring principle is drawn on the Newton Viscidity Theorem.
Calibration Procedure
There is a calibration function in the instrument test software. The standard viscosity liquid prepared by the National Standard Material Research Center is adopted.
1. When is calibration required:
1.1 The instrument is initially installed.
1.2 The instrument is moved, the computer system or the viscosity meter is changed or replaced.
1.3 After using the instrument for a period of time, it is found that the measured value of the instrument has obvious deviation.
☆Note: Before the instrument is calibrated, the horizontal position of the test movement should be adjusted: place the level meter on the test movement platform,