Cabinets with extractor system, designed for storing histological samples preserved in
formalin, or storing chemicals and solvents.
Construction features
-Electrogalvanized steel structure
-Three tray-type shelves, height-adjustable
-Leaf doors made of safety glass
-Control panel with soft-touch keypad forsetting the desired operating parameters
Extractor system
-Non-sparking extractor
-Alumina pellet filters for formaldehyde
-Manifold for connection to central extractor systems
Electrozinced sheet steel structure, doors and shelves, thickness 12/10, painted with grey (RAL 7035) antiacid epoxy powder.
Double door in 6/7 mm accident prevention stratified glass provided with closing on 3 points, lock and 180° opening in order
to make the shelves moving easier. Yellow danger signs in conformity with the regulations in force.
N. 3 extractable shelves with adjustable height, tray shaped in order to contain incidental liquid leaks. Capacity 50 Kg.
N. 1 AISI 304 stainless steel bottom tray, thickness 12/10, for the collection of incidental liquid leaks.
Palletized base.
N. 4 antiantistatic and antislip adjustable feet allow to get a perfect horizontal position.
Value of aspiration 0,50 m/s.
Visual and acoustic alarm for open door. The ventilation goes automatically at maximum power in order to safeguard the
safety of the operator.
Flow 300 m³/h
General ignition switch.
Emergency switch.
Start/stop switch for aspiration working
Fan speed regulation with graphic bar of percentage visualization.
Key fast adjustment of fan speed measured in percentage.
Visual and acoustic alarm in case of filters saturation.
Visual and acoustic alarm for open door.