KENZA MAX BioChemisTry PHOTOMETER. Compact & Convenient. Your complete working station for routine biochemistry.
One of the best-selling models in the world, many systems still running 10 years after their installation.
Integrated 9-positions incubator and 18 µL flow cell, very useful for medium series.
Storage hole for the flow cell : allow the reading on disposable cuvettes.
Open System.
Storage of calibration factors and reagent-blank values.
Storage of 30 QC values (on 2 levels for 30 analysis), means calculation, standard deviation, C.V.’s and printing of Levey-Jennings diagrams.
Built-in graphic printer : Different printing formats for results (with headlines, reference values, settings of parameter, etc…).
Printing of reaction curves (for kinetic methods) or calibration curves (multistandard methods).
Printing of report by ID and name of the patient (standard keyboard connection not supplied).
Key strengths:
• Easy to use photometer
• 120 programmable methods (Absorbance, End-Point, Kinetic, 2 Points Kinetic, Multistandard)
• Built-in printer
• 9 Positions incubators
• Fully Open system
• Robust
• Tutorial available for quick start of the analyser
Main features KENZA MAX BioChemisTry:
• Flow cell 18 µL
• Minimum volume per test : 500 µL
• Operating procedures : Absorbance, End-Point, Kinetic, 2 Points Kinetic, Multistandard
• Wavelengths: 8 interferential filters: 340, 380, 405, 505, 546, 578 et 630 nm, + 1 free position
• Measuring range : from -200 to +2500 mAbs
• Photometric accuracy : CV < 1µ to 2000 mAbs
• Photometric linearity : ± 1%
• Reproductibility : ± 1% de 0 à 2000 mAbs
• Incubator : 9 positions, from 20 to 40 °C