A sleek, simple design to carry with you. Frequent glucose monitoring is a breeze with the ELSA.
Tiny Blood Sample
Requires only 0.75µL.
Side-Insert Design
Easy to handle and prevents blood contact for excellent hygiene.
User-Friendly
Start monitoring quickly with this simple design.
Proprietary Strips
Enhanced size for improved grip, ensuring accuracy and precision.
How it works
Insert the strip into the socket of the meter. Calibration will occur automatically.
Touch a blood sample against the sample entry port of the strip.
Rotate and pull upwards to remove the test strip.
Specifications
Blood sample source - Capillary whole blood
Minimum sample volume - 0.75 µL
HCT range - 30 - 57%
Measuring range - 0.6 - 33.3 mmol/L (10 - 600 mg/dL)
Meter dimensions (L x W x H) - 95.0 x 43.8 x 13.0 mm (3.74 x 1.72 x 0.51 inch)
Display area (L x W) - 38.0 x 29.0 mm (1.49 x 1.14 inch)
Operating temperature - 10 - 40℃ (50 - 104°F)
Operating relative humidity - 10 - 90%
Meter storage conditions - -10 - 60°C (14 - 140°F)
Test strip storage conditions - At 4 - 30°C (39 - 86°F), and below 90% relative humidity
Accuracy - For glucose concentrations < 5.5 mmol/L (100 mg/dL),
95% of measured glucose values within ± 0.83 mmol/L (15 mg/dL) of reference
measurements.
For glucose concentrations ≥ 5.5 mmol/L (100 mg/dL),
95% of measured glucose values within ± 15% of reference measurements.
Precision - For glucose concentrations < 5.5 mmol/L (100 mg/dL), SD < 0.28 mmol/L (5 mg/dL)
For glucose concentrations ≥ 5.5 mmol/L (100 mg/dL), CV < 5%