Eliminates the need for manual coding and delivers maximum accuracy.
Large Display
Easy to view results.
Side-Insert Design
Easy to handle and prevents blood contact for excellent hygiene.
Tiny Blood Sample
Requires only 0.75µL.
Trusted Accuracy
Utilizes gold electrodes and a patented strip structure to provide accurate and reliable results.
How it works
Insert the strip into the socket of the meter. Calibration will occur automatically.
Touch a blood sample against the sample entry port of the strip.
After 5 seconds, the test result will appear.
Rotate and pull upwards to remove the test strip.
Specifications
Blood sample source - Capillary whole blood (finger, palm and forearm)
Minimum sample volume - 0.75 µL
HCT range - 30 - 57%
Measuring range - 0.6 - 33.3 mmol/L (10 - 600mg/dL)
Meter dimensions (L x W x H) - 98.0 x 46.0 x 17.5 mm
Display area (L x W) - 53.0 x 32.0 mm
Operating temperature - 10 - 40°C (50 - 104°F)
Operating relative humidity - 10 - 90%
Meter storage conditions - -10 - 60°C (14 - 140°F)
Test strip storage conditions - 4 - 30°C (39 - 86°F), < 90% relative humidity
Accuracy - 95% of the measured glucose values shall fall within either
Glucose concentrations < 5.5 mmol/L (100 mg/dL), difference within ± 0.83 mmol/L (15 mg/dL)
Glucose concentrations ≥ 5.5 mmol/L (100 mg/dL), bias within ± 15%
Precision - Glucose concentrations < 5.5 mmol/L (100 mg/dL), SD < 0.28 mmol/L (5 mg/dL)
Glucose concentrations ≥ 5.5 mmol/L (100 mg/dL), CV < 5%