Spec:
lnjection Mode - R+V,V+T,R+T,R
Volume Range - 0.1~9999.9ml,step by 0.1ml
Flow Rate - 2ml:0.1-100ml/h ;5ml:0.1-150ml/h;
10m1: 0.1-400m1/h; 20m1:0.1-600ml/h
30ml: 0.1-1000ml/h;50/60ml: 0.1-1800ml/h;
Step size 0.01mlh.
Time setting range - 00ho0min~99h59min,step by 1 min
Flow Rate Accuracy - ±2%
Purge Rate/Bolus Rate - 2ml : 100m1/h;5ml : 150ml/h;
10ml : 400ml/h; 20ml : 600ml/h;
30ml : 800ml/h;50ml : 800ml/h;
kVO Rate - Flow rate ≥10ml/h,KVO rate = 3ml/h;
Flow rate ≥1ml/h and <10ml/h,KVO rate = 1ml/h;
Flow Rate<1ml/h,KVO rate = flow rate
Alarm Functions - Occlusion, Near end of infusion,End of infusion,No
syringe,No holder,Low battery,Battery exhausted,A.C.
and D.C. off,Malfunction, etc.
Occlusion Pressure - 40±20kpa,80±20kpa,120±20kpa
Dimension - 297×147×106mm(W×DxH)
Weight - 1.4kg
Classification - Class ll,Type CF,IPX2
Compatible Syringe Size - 2ml,5ml,10ml,20ml , 30ml, 50/60ml
Standard Accessories - Pole clamp. Power adapter. User manual