The BI-DNDC is a deflection type refractometer that may be purchased for use in either batch or GPC/SEC mode. In batch mode, the specific refractive index increment, dn/dc, is determined. This value is required as a parameter in molecular weight measurements using light scattering. In GPC/SEC mode, dn/dc is already known, and the instrument is used as a concentration detector for GPC/SEC applications.
All major features are user programmable in order to achieve the maximum flexibility. The BI-DNDC can be set using the front panel controls and read using the LED readout or by an external computer. It includes a microprocessor, which allows user-friendly handling of all functions of the instrument, as well as refractive index and temperature data transfer to a personal computer via a standard serial interface.
Most importantly, warm-up time for the BI-DNDC is only one hour, much less than other models on the market.
Features at a glance
Fast instrument warm-up, ~ 1 hour!
Sensitivity, 2.5 x 109 ΔRIU
Autozero
Reference cell flush
9 measurement ranges
DNDCW Windows™ software
Active temperature control to 80 °C
Wavelength selection: 620, 535 or 470 nm
Determination of dn/dc
The dn/dc determination requires calibration of the batch mode instrument with a solvent/solute combination with known dn/dc, typically KCl in water. The calculated constant is not solvent dependent and may be used for all further measurements. In batch mode, the instrument comes with BI-DNDCW software for use with an external PC. Included with the BI-DNDC instrument is a syringe kit with PEEK™ and Teflon ® tubing, fittings and syringes.