• Flow ranges: up to 500 SLM
• Supply: 6 - 16 Vdc
• Analog (voltage) or Digital (I2C) linear output
• RoHS and REACH compliant
• Media Compatibility: dry gases only
For PMF4A00V-B (B for bi-direction calibration) Flow Rate = [(Vout - 3 V) / 2 V] x Full Scale Flow Rate For example, for PMF4A05-B full scale flow rate is 500 SLM. When Vout reads 2 V, the Flow Rate is: [(2 V - 3 V)/2V x 500 SLM] =-250 SLM (Reverse direction)
Order I²C Digital Output models by specifying “D” instead of “V” in part numbers above. Please contact Posifa or your local distributor to place an order.