The double terminal unit is a device
allowing, from a single medical terminal
unit, the duplication of the outlet
connection.
This device allows the connection, on a
single outlet, of two separate medical
devices requiring a gas / vacuum supply
(for example 2 flowmeters).
According to the selected location, this
device allows the orientation of the
double terminal unit and its blocking in
the best position.
Taking into consideration the maximum
flow available at its connection, ensure
that the cumulative flow need of the
connected devices respects the limits
specified in the ISO 9170 standard.
• Inlet pressure according to ISO 7396-1
• Gas: O2, AIR, Vacuum and CO2
• Inlet connection: AFNOR lockable probe
colored according to gas
• Outlet connection: AFNOR terminal unit
• Operating temperature:
from 0°C to + 40°C
• Body: Chromium plated brass
• Connecting hand wheel: plastic