Specifications
SpO2(*1)
Accuracy (*2)
70 % SpO2 ~ 100 % SpO2 : ±2% SpO2;
0% SpO2 ~ 69 % SpO2 : unspecified
Resolution 1 % SpO2
Plus rate
Accuracy (*3) ±2 bpm or ±2%, whichever is greater
Resolution 1 bpm
Battery 1.5V (AAA (LR03)size) alkaline batteries x 2
Dimensions Approx. W36 ×H32 ×D61 mm
*1The claims of SpO2 accuracy shall be supported by clinical study measurements taken over the full range. By artificial inducing, get the stable oxygen level to the range of 70 % to 100 % SpO2, compare the SpO2 values collected by the secondary standard pulse oximeter equipment and the tested equipment at the same time, to form paired data, which are used for the accuracy analysis.
There are 12 healthy volunteers (male: 6. female: 6; age: 18~45; skin color: black: 2, light: 8, white: 2) data in the clinical report.
*2Because pulse oximeter equipment measurements are statistically distributed, only about two-thirds of pulse oximeter equipment measurements can be expected to fall within ±Arms of the value measured by a CO-OXIMETER.
*3Patient simulator has been used to verify the pulse rate accuracy, it is stated as the root-mean-square difference between the PR measurement value and the value set by simulator.