The VPHP Testing Isolator was designed for analysis of the decontamination processes in the pharmaceutical field, by means of vaporised hydrogen peroxide (VPHP).
Air flow velocity probe
Temperature and humidity sensor
Analysis system for H2O2
Integrated sensor for detecting the concentration of hydrogen peroxide, temperature and humidity
H2O2 trend detection sensor (maximum level)
Low concentration H2O2 detection sensor (minimum level)
Photocatalyst
Dehumidifier
Speedy Glove system
Glove spreaders
Internal shelves
In particular, the isolator is used for:
Implementation and control of decontamination cycles by means of VPHP for the evaluation of H2O2 residues inside the chamber and its exposed materials once completed.
Monitoring and recording H2O2 concentrations during the entire decontamination cycle and monitoring extremely low concentrations (less than 3 ppb) in the final phases of decontamination by means of a high sensitivity sensor.
Carrying out desorption tests on materials inside the chamber.
The isolator works in positive pressure with respect to the laboratory to ensure chamber sterility.
The isolator is composed of a pre-chamber, used for the inlet of materials, and of a main chamber for implementing analyses.
The isolator is equipped with a VPHP generator for the decontamination of the work areas.