This device is a separated SpO2 simulator which features in small volume and light weight. It can perform a series of tests by the simulation means, to provide an accurate reference for the Pulse Oximeter.
As different manufacturers may use different R-curve, current popular R curves (such as BCI, Masimo, Nellcor, etc.) are embedded into the simulator in advance.
Function
1)Simulation of SpO2 and pulse rate.
2)Support SpO2 and pulse rate simulation under 60 Hz/50 Hz interference and light interference.
3)Adjustable perfusion index (PI), support SpO2 and pulse rate simulation under different PI.
4)The reaction time of SpO2, PR and cardiac arrest can be tested.
5)Built-in 10 compatible R curves (such as BCI, Masimo, Nellcor, etc.), meet the test requirements (R curve can be customized).
6)Preset 24 patient states, such as weak pulse, bradycardia, hypoxia, infant, tachycardia, the aged, obesity, etc.
Main parameters
1)SpO2
Range: 35 % ~ 100 %
Resolution: 1 %
Accuracy: when SpO2 is 75 % ~ 100 %, the error is ± 2 % or ± the device's accuracy, whichever is greater; when SpO2 is 74 % ~ 50 %, the error is ± 3 % or ± the device's accuracy, whichever is greater; no definition when SpO2 is less than 50 %.
2)Pulse rate
Range: 20 bpm ~ 250 bpm
Resolution: 5 bpm
Accuracy: 1 % ± 1 bpm
3)Low perfusion
Range: 0.000 % ~ 20.000 %