Introduction
The device adopts sidestream method to measure the parameter related to ETCO2 and AWRR. It can detect pulmonary ventilation function and reflect the circulation and pulmonary blood flow and indirectly reflect the alveolar ventilation. Meanwhile, it can also detect the blood oxygen saturation and pulse.
The application population is adult patient. It is applicable for use in emergency department, ICU, operating room, respiratory department and other occasions. It is connected to the nose end of the patient through the sampling tube (applied part) .
Features
1)End-tidal CO2 measurement function;
2)Respiration rate measurement function;
3)SpO2 measurement function;
4)Pulse rate measurement function;
5)Optional three units;
6)With audible and visual alarm and silencing functions, the alarm limit can be set;
7)Over-limit alarm function for EtCO2, Respiration rate, SpO2 and Pulse rate;
8)Low battery alarm function;
9)With the prompt function of “Checking Filterline”;
10)User can calibrate zero by themselves.
11)With atmospheric pressure and temperature compensation functions
12)Rechargeable lithium battery, up to 8-hour user time;
Performance
End-tidal CO2 measurement:
1)Measurement range: 0 mmHg ~ 150mmHg
2)Resolution: 1mmHg
3)Accuracy:
0 mmHg ~ 40 mmHg:±2 mmHg;
41 mmHg ~ 70 mmHg:±5% of readings;
71 mmHg ~ 100 mmHg:±8% of readings;
101 mmHg ~ 150 mmHg:±10% of readings.
Respiration rate measurement:
1)Measurement range: 2rpm~120rbpm;
2)Resolution: 1rpm;
3)Accuracy: ±1rpm.
SpO2 measurement:
1)Measurement range: 30 % ~100 %;
2)Resolution: 1 %;
3)Accuracy:
70 % ~ 100 %:2 %;
30 % ~ 69 %:undefinition.