Criticare’s Poet® IQ Anesthetic Gas Analyzer pairs with the nCompass™ Patient Monitor, to provide a unique combination of leading edge vital signs technology and anesthesia gas monitoring in a compact, modular system.
The Poet® IQ automatically identifies and quantifies inspired and expired O2, CO2, N2O, and five anesthetic agents. The system’s reliable performance, ease of use, flexible design, and affordable cost make it the ideal monitoring solution for anesthesia applications in hospitals and surgical centers.
Features
State-of-the-art non-dispersive infrared (NDIR) technology identifies and measures five anesthetic agent gases: Halothane, Enflurane, Isoflurane, Desflurane and Sevoflurane
Automatic or manual modes for agent detection
Adjustable flow rates: 100, 150 and 200 ml/min
Mixed agent identification
Accuracy not affected by alcohol or ketones
Fast warm-up time ensures full accuracy within minutes
Specifications
GAS MONITORING
Method: - Sidestream; Non-dispersive infrared (NDIR)
Identified Gases: Halothane, Enflurane, Isoflurane, Desflurane, Sevoflurane, O2, CO2, N2O
Concentration Units: Vol%, Torr, kPa, mmHg
Flow Rates: 100, 150 or 200 ml/min
AGENT DETECTION
Measurement Range: - Halothane: 0 to 10%, Enflurane: 0 to 10%, Isoflurane: 0 to 10%, Desflurane: 0 to 20%, Sevoflurane: 0 to 10%, CO2: 0 to 12.5%, N2O: 0 to 99%, O2: 0 to 100%
Measurement Accuracy: Agents: ±0.1% abs. + 4% of reading, CO2: ±0.2% abs. or 4% of reading, N2O: ±1.5% abs. + 4% of reading, O2: ±3 vol% (0 – 90%), ±4 vol% (91 – 99%)
Time to Detect Agent: < 15 seconds @ 200 ml/min
Agent Detection Resolution: 0.1 vol%
Mixed Gas Threshold: - 0.2 vol% +10% of total concentration