NEO-BIL PLUS is an advanced and innovative version of traditional bilirubinometers. It is to analyze the ‘total bilirubin’ through a microcapillary tube in order to follow the new born icterus course. More than 20 years of experience in this specific field and that of physics-chemestry have led us to this achievement. Thanks to the use of the most advanced technology, frequent adjustments and continuous standard resetting are over. A precise sample measurement of both bilirubin and interferentials on the same point of the capillary is carried-out by a special reading system. The reading is carried out even on a small quantity of serum. No reagents are required. NEO BIL Plus is equipped with a built in printer for the results sequential printing.
TECHNICAL FEATURES
Intended use - measurement of total bilirubin
Sample - centrifuged whole blood
Reading cuvette - haeparinized haematocrit capillary
Measure unit - mg/dl or µmol/l
Measurement range - 0/30 mg/dl or 0/510 µmol/l
Measure system - photometric
Sample reading - on a single-point of the capillary
Zero and checkings - completely automated
Interferentials - automatically compensated
Reading time - 3s approx. even on highly haemolized samples
Reading inaccuracy - 3% in the range 5-30 mg/dl or 85-510 umol/l
Detector - silicon photodiode
Programming - by membrane keypad
Results - on LCD and printer
Keypad - 17 buttons
Display - 2 lines with 16 columns
Printer - 20 columns with termal paper
Data storage - last 1000 results
Interface - USB port
Other functions - built-in calendar and timer
Power requirements - 230/115Vac, 50-60Hz, 25W