The Next Generation Infant Incubator & Radiant Warmer Analyzer
Tests to IEC 60601-2-19, 2-20 and 2-21
Follows both Edition 1 and Edition 2 of Standard
Built on the Vision-Pad platform to give you greater flexibility and scalability
Wireless control of tests within incubator
Changes to test can be made without disturbing heat environment
Measures all parameters automatically following user-defined protocols
Results can be merged with Electrical Safety test results
vPad-IN is a next generation infant incubator and radiant warmer testing system that meets the requirements of the IEC Standards for Incubators (IEC 60601-2-19), Transport Incubators (IEC 60601-2-20) and Radiant Warmers (IEC 60601-2-21).
vPad-IN has all the functionality and scalability that you have come to expect from Vision-Pad Technology products. Parameters are sensed inside the incubator environment and results are transmitted wirelessly to a 10.1″ Android tablet for analysis and display. Test parameters and test protocols can be modified without disturbing the equilibrium of the heated environment, thereby minimizing test times.
Compatible with closed, forced air convection incubators, and open, radiant infant warmers.
Simultaneously measures humidity, airflow, sound, and 6 different temperatures
Tests are carried out in accordance with IEC 60601-2-19, IEC 60601-2-20, and IEC 60601-2-21
All sensors are independently movable, to be placed in positions required by the Standard
Sound level measurements may be taken outside the incubator for external alarm test
Results can be merged with vPad Electrical Safety test results
Optional precision controlled heater can accurately test body temperature sensors