GENERAL DESCRIPTION
Terminal units for Compressed Air and Nitrogen to feed surgical tools have been projected in accordance with ISO 9170-1 and to grant an high flow rate with a low load loss. The gas flow is activated by the specific probe. The double valve system allows the maintenance of the external valve without stopping the gas supply in the main pipeline.
MAIN FEATURES
Available base blocks: with soldering coupler or with hose connection
Available probes: with 90° hose connection or G1/4 female connection
Regulatory Standards: ISO 9170-1
CE marked products as medical devices in accordance to 93/42/EEC Directive
Degreased in accordance with ISO 15001
These outlets are supplied to be installed on bedhead units by a proper kit or for the wall installation in external or wall.
The vacuum level is adjusted by the proper regulation valve and maintained stabilized by the intermittent works of the pumps. The transducer read and shows continuously the generated vacuum level.
Terminal units for Compressed Air and Nitrogen to feed surgical tools have been projectedin accordance with ISO 9170-1 and to grant an high flow rate with a low load loss.
The gas flow is activated by the specific probe.The double valve system allows the maintenance of the external valve without stopping the gas supply in the main pipeline.
These outlets are supplied to be installed on bedhead units by a proper kit or for the wall installation in external or wall.
Technical Data
SUITABLE FOR FOLLOWING GASES: N2-800 . Air-800
MAX INLET PRESSURE: 20 bar
OPERATING PRESSURE: 5.6 * 12 bar
STORAGE TEMPERATURE: -20 + + 60 °C