Optional with I.S.E module (Na, K, Cl, Li)
Test modes: Routine, Batch, Emergencies (STAT), Profiles
Up to 500 tests in memory+ unlimited relation tests
Sample: Serum, Urine, Plasma, CSF
Sample & Calibrator pre-dilution
Cooled reagent tray
Cooled calibrator & control tray
Non disposable optical glass cuvettes
Solid state photometer with 14 channels 340-800nm
Clot sensor
Crash sensor
QC program
Internal barcode reader for sample and reagents
Bidirectional LIS
No existing water and waste lines necessary
Photometer
Solid state photometer: 14 channel + 1 ref. channel
Wavelengths: 340, 380, 405, 436, 478, 510, 546, 578, 600, 630, 660, 700, 740, 800 nm
Absorbance range: 0.000 – 3.000 Abs
Sensitivity: ± 0.001 Abs
Accuracy: ± 1% from 0 to 2.000 O.D
Tungsten halogen lamp with dichroic reflector
Lamp Life time approximately 2000 hours
Pippetting arms / Dilutor
1x Sample pipetting arm
2x Reagents pipetting arms
3x Long life ceramic dilutors
Dilutor precision: ± 0.1 %
Dilutor resolution ± 0.1 μL for serum
± 1.0 μL for reagents
Power Requirements, Weight and Dimensions
Power supply: 100 / 240 VAC, 50-60 Hz, 550 Watt