InnovaStar® is an automatic compact POCT analyzer for the daily practice that delivers result quality comparable to measurements on laboratory analyzers.
Easy to use
No special knowledge and extended training is required to operate the system. Testing is made easy with unit-dose reagents and precalibrated tests. The system is calibrated automatically with parameter cards which come with every test package.
Precise
High-quality reagents from DiaSys, freedom from interference, and a photometer detection unit with outstanding linearity are the guarantee for highest possible precision.
Technical specifications
Barcode Identification - Automatic barcode scan for reagents; optional barcode scanner for samples
Calibration - Pre-calibrated tests
Data Storage - 50 results
Measurement Principle - Photometric / Turbidimetric
Method Storage - Up to 19 different methods or lots
EDP Interface - RS232 connection; USB connection via adapter
Dimensions - 20 cm (W) x 15 cm (H) x 17 cm (D)
Parameter HbA1c IS
Measuring Range - 4.9-16%DCCT
(unitsselectable) - 30 -150 mmol/mol
Precision (CV%) - 0.985% at 5.58% DCCT
(in series, n=20)
Samplespecimen - WholeBlood
Sample Volume - Détermination of ail analytes
Measuring Time - <5 min
Shelf Life - 18 months
Weight - 4 kg