Control Type - Microprocessor Controlled Unit
Power Requirements - 220VAC, 50 Hz
Heater Capacity - 500W (Ceramic heater)
Canopy mobility range - Horizontal rotation 3200(1600 both clockwise and counterclockwise Radiation range +300
Temperature control - Servo (Skin) Mode/ Manual/ Preheating mode
Skin temperature -
Setting range: 34.0~38.00C
Accuracy: + 0.30
Display range: 20.0~42.00C
Heater Output -
Setting range: 0~100%
(in 5% increments)
Indication range: 0~100%
(in 1% increments)
Timer
APGAR mode /CPR
Mode switchable
APGAR Beep: 1,3 (option), 5,10 min
Examination lighting
Power: 6W LED (2W * 3)
LED lamp Adjustable illumination intensity: 0~1500lux
LCD display
10” LCD touch screen
Pixel: 800 * 600
Bed tray
Inclination angle
(back and forth : +130
Mattress surface: 500 * 700 * 25mm
Mattress surface to ground height (standard): 90cm
Guard door: damping system
Cable access hole: 3
Dimensions
84(W) x 110(D) * 168(H)
Cm (standard)
Weight - 110kg
Alarms Silence - Hand free
Height adjustment - Up to 30 cm
Weight Scale
Weight range - 300 g to 8000 g
Weight display resolution - 1 g
Weight accuracy - +10 g
Drawable X-Ray Cassette Tray
Dimension - 33 * 35 cm
Drawer
Weight limit - 5 kg
Pulse Oximeter Unit
Display range - 1~100%SpO2
Display accuracy - +2% (70~99%)
Pulse rate display range - 30~240bmp
Pulse rate display accuracy - 1% + 1 bpm
Resuscitation
Manometer - Pressure indication: -2~10 KPa (-20~100cmH2O)
Resusci-flow unit - PMAX setting range
<6kPa (60cmH2O)
Suction Unit
Negative pressure range - Suction pressure control range:
-24~0kpa(-180~0 mmHg)
Air pumping efficiency - 0~15L/min