Input power 120 VAC, 50 to 60 Hz.
24” flexible stainless steel braided pigtails. Vertical crossover
and staggered styles include 36” pigtails for half of the
cylinders. All pigtails include a check valve.
36” copper pigtails with check valve supplied with oxygen units
CGA connections with integral check valves at each header station.
Unit of measure switching (psi, kPa, BAR).
Dual line pressure regulators
Special header configurations available upon request (Ushaped,
L-shaped, etc.). (Dimensional sketch of installation
required).
Built for expansion by adding header extensions.
Cabinet weight 70 lbs.
May be converted from high pressure cylinder use to use with low or medium pressure liquid portable bulk vessels.
Line pressure sensor may be mounted inside the cabinet or remotely located to eliminate the need for a high/low pressure switch for master alarm operation.
SPECIFICATIONS
The NFPA 99 compliant digital, fully automatic manifold requires no manual resetting of valves or levers. The unit switches from “Bank in Use” to “Reserve” bank without fluctuation in line delivery pressure. Simultaneously, the “Reserve in Use” alarm is triggered by the manifolds microprocessor. The manifold will continue to provide gas, in the event of a power failure, until both banks are depleted.
After the switchover, the “Reserve” bank will then become the “Bank in Use”. The manifold microprocessor shall also trigger the “High Line Pressure” and “Low Line Pressure” alarms without the need for additional pressure switches or transducers.