Normally, image plate scanners only have a fixed laser beam diameter for capturing data from image plates. With the new DualFoc technology from DÜRR MEDICAL, new optics make it possible for the first time to set two different laser beam diameters within the same scanner.
What is the advantage?
Unlike all other devices on the market, the CR 35 VETwin is the only scanner fully utilizing the properties of both intraoral and extraoral image plates. Being limited to just one fixed laser beam diameter for scanning the image plate means you are forced to find a compromise between high resolution and low signal noise. Through the use of two different foci, it becomes possible for the first time, to define perfect parameters both for high-resolution intraoral plates and for the plates generally used for full-body images.
The operating principle
High resolution blue image plates for dental radiography are scanned automatically with a laser beam focused to 12.5 µm (25 lp/mm). This way images of unrivaled quality are created.
When conventional image plates are used for non-dental radiography, the laser beam diameter is focused to 50 µm (10 lp/mm), which guarantees brilliant low-noise images with high contrast for best diagnosis.
Impressive functions in a single device
4,3" HD-Display
- High resolution(800 x 480 px)
- Full color display
- Touch functionality
- With own Stylus
Built-in mini-PC
ln combination with the built-in mini-PC, the high resolution full-color display, integrated into the housing, allows to operate the device fully independently.