Central heat disinfection system with ultrafiltration and storage tank for heat disinfection of the permeate ring and connection hose through into the dialysis machine.
By continuous heat disinfection of the distribution ring (> 80 °C) and at an exposure time of >30 minutes vegetative cells of microorganisms are destroyed safely and prevents biofilm formation.
Permanent ultrafiltration during dialysis ensures highest permeate quality. The currently required microbiological permeate quality will be surpassed by this technology. Even future demands for ultrapure permeate will be met with this system – which confirm the long-standing experience of many satisfied users.
480 litre isolated hot water tank
Integrated pressure pump
Individual timing avoids electrical peak loads
Volume control – filling level can be individually adjusted between 200 litre and 480 litre
Loop line to the dialysis unit
Temperature control – selectable and controllable temperature of the tank volume
Conformity declared with directive 93/42/EEC
Patented procedure of Ultrafiltration of permeate
High lifetime of the ultra filter by cross-flow operation
3-step permeate pressure valve for fully automated filling of the tank at permeate oversupply
Master role in connection with the Reverse Osmosis System