ACCUGENCE ® LITE Multi-Monitoring System is a simple alternative to ACCUGENCE PM900 that let you do all the main things you’d expect from a basic multi-monitoring monitoring system with less cost compared to PM900.
Advanced Features:
4 in 1 Multi-Function
Underdose Detection
New Enzyme Technology
Small blood sample volume
Wide operating temperature
Comprehensive quality control
Reliable result
Clinical proven performance
Full Compliance ISO 15197: 2013 / EN ISO 15197:2015
Technical Specifications
Parameter
Blood Glucose, Blood β-Ketone, and Blood Uric Acid
Measurement Range
Blood Glucose: 0.6 - 33.3 mmol/L (10 - 600 mg/dL)
Blood β-Ketone: 0.0 - 8.0 mmol/L
Uric Acid: 3.0 - 20.0 mg/dL (179 - 1190 μmol/L)
Hematocrit Range
Blood Glucose and β-Ketone: 10% - 70 %
Uric Acid: 25% - 60%
Sample
When testing β-Ketone, Uric Acid or Blood glucose with Glucose Dehydrogenase
FAD-Dependent, use fresh capillary whole blood and venous Blood samples;
When testing Blood glucose with Glucose Oxidase: use fresh capillary whole blood
Units of Measure
Blood Glucose:
The meter is preset to either millimole per liter (mmol/L) or milligrams per
deciliter (mg/dL) depending on the standard of your country.
Blood β-Ketone:The meter is preset to millimole per liter (mmol/L)
Blood Uric Acid:The meter is preset to either micromoles per liter (μmol/L) or
milligrams per deciliter (mg/dL) depending on the standard of your country.
Automatic Shutoff
2 minutes
Meter Size
79 mm × 50 mm × 14.5 mm
On/Off Source
One CR 2032 3.0V coin cell batteries
Battery Life
Around 500 tests
Display Size
30 mm × 32 mm
Operating Temperature
Glucose and Ketone: 5 - 45 ºC (41 - 113ºF)