Optimised for MCS
14 Pre-Set pulse widths from 50 ns to 1 ms
External trigger capability
Adjust pulse width and repetition rate with input trigger pulse
Operation in CW mode
Spectrally Purified Output
Compact plug-and-play design
Product Description
The VPLED series of LEDs provide variable width pulses from ns to ms, as well as CW operation with powers in the mW range.
They are optimised for long photoluminescence lifetime measurements using the Multi-Channel Scaling (MCS) detection method. The width of the excitation pulses is easily controlled from the VPLED allowing quick optimisation of the measurement conditions.
The VPLED diodes are compact, maintenance free, and user-friendly. Although they are designed to match Edinburgh Instruments spectrometers, they can be incorporated into any experiment thanks to their external trigger capability.
Technical Specifications
External Trigger Width Mode - Pulse width and repetition rate defined by external trigger input
Max. duty cycle 50%
CW Mode - Yes
Bias Supply - 15 Vdc +/- 5%, 15W (2.1mm DC jack)
Trigger Output - SMA, NIM Standard
Interlock Input - Hirose HR10A-7P-4P(73), (Link pin1 and pin 2 to ground – interlock healthy)
Trigger Input - Hirose HR10A-7P-4P(73), (Signal pin 4 and ground pin 3)
Trigger Input Signal - TTL > 50 ns pulse. Trigger on rising edge.
0.0 V < Low level < 0.5 V, 2.5 V < High level < 5 V
Key Switch - Yes
Spectral Conditioning - Built-in filter to minimize out-of-band emission (no spectral filtering needed)
Physical Dimensions - Overall: 168 mm length x 64 mm x 64 mm
Collimator tube: φ 30 mm x 38 mm
Tapped Holes for Stud Mount - 2 x M6
Weight - 750 g