Temperature data-logger ECOLOG TH2
humidity2-channel

temperature data-logger
temperature data-logger
temperature data-logger
Add to favorites
Compare this product
 

Characteristics

Data type
temperature, humidity
Other characteristics
2-channel

Description

2-channel temperature/humidity logger, Measuring range: -35 °C..70 °C or 110 °C with high TH probe; Operating range ECOLOG housing -35 °C..+55 °C and 0..100% RH, Memory capacity: 32'000 values per channel, Connections: RS232, 2 digital inputs, alarm output, Validation Certificate (validation points -0.4, +73.3 °C and 5% RH, 88.4% RH @ 23 °C). Characteristics Product type:Datalogger Logger/module type:Logger, not networkable Probe type:Combi-probe T/RH, external Typical application area:Freezer, refrigerator, room, climate room Measurement parameters:Temperature / Humidity Measurement range temperature:-35 °C..+70 °C / 110 °C Measurement range humiditiy:0% RH..100% RH Measurement accuracy temperature:±0.2 °C [+0.1 °C..+30.0 °C] / ±0.3 °C [-24.9 °C..±0.0 °C; +30.1 °C..+70.0 °C] / ±0.4 °C [-50.0 °C..-25.0 °C; +70.1 °C..+100.0 °C] / ±0.7 °C [+100.1 °C..+140.0 °C] Measurement accuracy humidity:±1.5% RH [+23 °C] Usage:Multiple Use Alarm:Alarm thresholds Connectivity:DB 15 Life time:10 Years Memory capacity:64'000 measurement values per logger Dimensions:110 x 35 x 85 mm (4.3 x 1.4 x 3.4) Material:ABS plastic material Certificates:Validation certificate

Catalogs

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.