Electronical Leakagetester
Power supply: - 230 V / 50 Hz
Fuses: - 2 x 1,25 A
Supervision mark: - VDE 0750 (part 1/05.02)
Width: - 280 mm
Height: - 150 mm
Depth: - 300 mm
The LEAKAGE:MASTER does not only serve for the examination of the endoscope on tightness, but finds also use already from prepurifying up to the end of the disinfection procedure and protects your endoscope against liquid break-down by permanent, electronically controlled counter-pressure.
During the manual precleaning and disinfection the Leakage:Masters avoids an ingress of water into the endoscope by permanent air feed charge via the ETOvalve.
The Leakage: Master is working with a counterpressure of approx. 200 mbar. Any possible leak is indicated visual and optical. The counterpressure will be sustained until the endoscope will be taken out of the fluid.
Therefore the Leakage: Master prevent high expenses for endoscope repair.
Suitable for all common, flexible endoscopes
Exact readout by display
Detects micro-perforations in less than 2 minutes
Optical and acoustical alarm signal
Leakage surveillance during the whole endoscope reprocessing