The VLF R Series from Envair guarantees superior product protection by offering a positive pressure particle-free working area achieved by initially pulling air through a G4 pre-filter (EN779:2012) and then pushing air through a High Efficiency Particulate Air (HEPA) filter type H-14 (BS: EN: 1822).
The VLF R Series is designed for NHS hospital pharmacies for CIVAS (Central Intravenous Additive Service), TPN compounding and bagging, preparation of IV solutions and eye-drops, ampoule filling, etc.
It has been adapted and customised for integration of paperless medication workflow solutions such as BD CATO™, and automated compounders such as the Baxter ExactaMix® 2400.
The VLF R Series design is ideal when freedom of movement into and in the working area is paramount and when sited in a Grade B cleanroom.
DESIGNED WITH CLEANROOMS IN MIND
The VLF R Series has been designed with the application in mind. All materials are specified to be non-shedding and resistant to common cleaning chemicals, with low noise and low energy consumption as standard with EC/DC fans and LED lighting (providing <750 lux in the work zone). This provides the most comfort and safety possible.
The VLF R Series is typically built to a clients URS, often including polished 316L stainless steel internally and externally, vari-height electrically operated height adjustable support frame, CCTV camera, sealed touchscreen monitor in rear wall of main chamber, power and data connection.