*Temperature range +4º C to +60ºC .).
*Humidity control: Working range from 10% to 90% R.H. (+/-3%), in a temperature range +19ºC to +40ºC.
*S4000 controller in EICHS test models, with graphical representation of data.
4.3" backlit TFT touch screen for temperature and/or humidity control with graphical representation of temperature and/or humidity curve.
*Precision of +/- 0.1ºC on screen. Homogeneity of + /-1.0 ºC inside the chamber at 37ºC. Stability of +/-0.5ºC inside the chamber at 37ºC.
*Refrigeration system controlled by solenoid valves.
*Safety thermostat, protects samples against high temperatures.
*The microprocessor control parameters control the temperature through a Pt100 probe, with a resolution of +/- 0.1ºC. Control system with battery backup, with automatic recharge, duration up to 48h.
*Electronic probe (4-20mA) with an accuracy of +/- 2% RH (optional 1.0 % R.H.).
*Generator of humidity by ultrasound introducing in the chamber microscopic drops of water at room temperature (optionally electrodes).
*Cooling condensation drying system.
*Internal door of thermal glass, with hermetic closing.
*Opaque external door, with opening aids and large handle.
*Optional: double glazed external door with large handle.
*Microprocessor for control parameters with PID system.
*Independent alarms, visual and audible, for maximum and minimum temperatures, with NiCd backup battery, of more than 48 h duration (72h optional).
*System settings protected by numeric password.