Extemal surface of ventilator (including casing,power cord, high pressure tube)
Wet the soft clean cloth in the cleaning agent (Soap water with PH ranges from 7.0 to 10.5, Concentration of 5%),
Thoroughly wipe the surface of the sample (Replace a clean soft cloth after each wipe)until there are no visible pollutants to the naked eye
Expiratory valve assembly
Including expiratory valve, mebrane
And pressure sampling pipe
Répétitive breathing line
Including liquid accumulation bottle
1. Rinse with running water to remove surface contaminants;
2. Soak the contact part between the sample and the human body in the cleaning agent (Soap water with PH ranges from
7.0 to 10.5, Concentration of 5%)for 5 min ;
3. Wipe the surface of the sample until there are no visible pollutants to the naked eye ;
4. Rinse the sample in purified water for 3 times, 1 min each time, and remove the residual cleaning agent on the surface ;
5. Use a dry and absorbent soft cloth to wipe off residual water stains on the sample, and place it in a dry place to air dry ;
6. Autoclaving, 134°C/188.4kPa ~ 210.7kPa/ 15mins ;
7. After disinfection, check if there is any steam residue inside the component. If there is, placeit in a dry place and let it dry without moisture before reuse.