Complete range of tests for the control of the sterilization process.
Bowie & Dick Test, Helix Test and chemical indicators, check the correct penetration of the steam inside the sterilization load, thus ensuring an adequate operating performance.
BOWIE & DICK TEST
FUNCTION
Check the correct steam penetration in the porous loads such as gauze, bandages, gowns and fabrics.
FREQUENCY
Daily
At the beginning of each shift
In the event of displacement of the autoclave
After any repair or maintenance intervention
CYCLE TYPE
B type cycle at 134°C for 3.5 minutes
HELIX TEST
FUNCTION
Check the correct steam penetration in the close lumen, such as turbines and contra-angles.
FREQUENCY
Daily
At the beginning of each shift
In the event of displacement of the autoclave
After any repair or maintenance intervention
CYCLE TYPE
B type cycle at 134°C for 3.5 minutes
B type cycle at 121°C for 15 minutes
CHEMICAL INDICATORS
FUNCTION
Check the critical variables such as time, temperature and presence of steam.
FREQUENCY
Single indicator for packaging for each sterilization load
CYCLE TYPE
B type cycle at 134°C for 3.5 minutes
B type cycle at 121°C for 15 minutes