Neonatal Non-Invasive Blood Pressure [135mmHg] Excess-Pressure Protection.
Adjustable Neonatal Breath Pause Alarm Time (15 - 20 s)
Monitor Chart Trends and Reviews Up to 72 hours.
Data Recoverable and Charts Can be Replayed.
Oxygen Level Can be Monitored Seperately with Functions of Setting High and Low Limits of Oxygen Levels.
Data Deposits and Review Functions Enable Caretakers for Further Examination.
Light and Portable for External Record and Connection to Central Networks
DC/AC Applicable with Internal Batteries for Up to 1.5 h Independent Power Supply.
Noncontact Breathing Head Mask Is Tailored and Designed for Newborns
Hear-Rate Modes
Lead Input:3/5 (3 for most cases)
3-Lead Mode:Ⅰ、Ⅱ、Ⅲ
5-Lead Mode:Ⅰ、Ⅱ、Ⅲ、AVR、AVL、AVF、V
Gain:2.5mm/mV,5.0 mm/mV,10 mm/mV,
20 mm/mV
Wave Speed:12.5mm/s,25 mm/s,50 mm/s,
Accuracy:±2% or ±1bpm
Resolution:1bpm
Leakage Current:<10μA
Protection:Separation Tolerance 4000V,50/60Hz voltage with anti-tremor protectin
Electrocardio Tabs:1mV±5%
Test Range:24~280bpm(b/min)
Alarm Range:100~180bpm(b/min)
ST Interval Analysis:Yes
Arrhythmia:Yes
Arrhythmia Alarm/Review:Yes
Alarm:multi-angle sound & light alarms with repetitions
Blood O2 Saturation Specification
Resolution:1%
Accuracy:80~100%±2%
0%~79% nondescriptable
Alarm Range:0~100% continuous
Resolution:1bpm
Accuracy:±2bpm
Alarm:multi-angle sound & light
Breathe Mode
Mode:RA-LL impedence
Resolution:1rpm
Accuracy:±2rpm
Asphyxiation Detection and Alarm:15~20 s (adjustable by seconds)
Blood Pressure Type
Measurement Method:
automatic oscillation
Work Mode: manual/automatic/continuous
Auto-Measure Time:
1~960 min (adjustable)
Measurement Unit:mmHg/KPa